Virtual Test Equipment Interface (9407)
The spectral information for the 20 kHz/Div and 50 kHz/Div spans might have a small frequency offset
Release Date: 1999/03/31
Affected Model(s): 9407
Affected Product(s): LVDAM-COM 1.01
Brief Description of the Problem:
When using the 9407 Virtual Test Equipment Interface with the Spectrum Analyzer, the spectral information for the 20 kHz/Div and 50 kHz/Div spans might have a small frequency offset.
Full Description of the Problem:
The central frequencies of the filters in the Spectrum Analyzer PCB inside the 9407 module need to be calibrated. A subroutine is missing in the initialization of the module and in the autocalibration of the Spectrum Analyzer program.
How to Fix:
Replace files TAIVTH00.EXE and TAIVTH01.EXE in folder \\LVDAM\\COM1_01\\.
Those files are available by downloading and running the self-extracting file PF1021.EXE.
Download pf1021.exe
Unable to open the Data Table window.
Release Date: 04-05-06
Affected Model(s): 9481, 9407, 9062
Affected Product(s): LVSIM-DCOM, LVDAM-COM, LVDAM-EMS
Brief Description of the Problem:
Unable to open the Data Table window.
Full Description of the Problem:
The Data Table window cannot be opened in any one of the aforementioned Lab-Volt software products. In certain situations, this may cause the software from which the attempt to open the Data Table window is performed to close.
How to Fix:
Download and run the self extracting files which will install the files needed to open the Data Table window without any problem.
Download pf1049.exe